Instruction Manual
1. Features
The AT210 wireless digital thermometer can be used for remote
temperature monitoring. It consists of two units: AT210-A (thermometer
unit) and AT210-B (remote monitor unit). The AT210-A connects to the
temperature sensor and displays measured value. It also sends the
measured temperature reading to the AT210-B and allow remote
monitoring and remote parameter setting through wireless connections.
The AT210-B can communicate with up to 10 units of AT210-A. The
AT210-A and AT210-B thermometers have two alarm settings for a built-
in buzzer. One can be used for high limit alarm and the other can be
used for low fuel (limit) alarm. This thermometer can also store the
highest temperature reading with a time stamp during the temperature
monitoring process. The AT210-A has a 12VDC 2.5mm power output
that is synchronized with the build-in alarm buzzer, which can be used
to drive an external buzzer or other devices. Both units are powered by
12VDC through a 2.1mm AC adapter for continuous operation. It can
also be powered by a car battery directly.
2. Specifications
Power input: 12VDC (2.1mm OD connector)
AC adaptor: 100-240V, 50/60Hz input. 12 VDC, 1 Amp output
Power consumption: <0.5W
Sampling rate: 2 samples/second
Accuracy: 0.2% full scale
Thermometer reading range: -320 ~ 2300° F (-200 ~ 1300° C)
External buzzer output (on AT210-A): 12VDC (2.5mm OD connector), 1
Amp, synchronized with the internal buzzer
Internal buzzer: two alarms (high limit or low limit alarm)
Dimension: 2.8 x 3.5 x 1.2" (70 x 90 x 30 mm)
3. Front panel
1
2
4
5
Figure 1. Front panel of AT210-A and AT210-B.
① Digital display window.
② Alarm indicator: blinking when alarm is triggered.
③ Peak value indicator: stay solid on when peak value is displayed;
blinking when the time stamp of the peak value is displayed.
④ Set key: enter code; confirm input value.
⑤ Shift key: shift digit; mute the alarm buzzer.
2020.09
AT210 Wireless Digital Thermometer
Version 2.0 (Sep, 2020)
3
7
6
⑥ Down key: change digit value; move to the previous parameter;
display peak temperature and the time when peak temperature is
reached/ reset peak values.
⑦ Up key: change digit value / move to next parameter / change
brightness.
4. Connectors on the meter
The connectors on the meter are shown in Figure 2. Connecting the 12V
DC power adapter to the connector 1 from wall outlet. Connecting the
external buzzer to connector 2. The polarity for this socket is center pin
positive (+), outer collar negative (−). Connect the K thermocouple to
terminal 3. Please note that thermocouple connector also has polarity.
The wide blade should go to the wide slot.
Figure 2. Bottom view of AT210-A (left, thermometer unit and AT210-B
(right, remote monitor unit).
① Power Input - 12VDC power supply input (2.1 mm O.D.).
② Power Output - 12VDC output (2.5 mm O.D.).
③ Probe Input - temperature probe input.
5. Basic functions for temperature monitoring
During the normal temperature monitoring mode, you can mute the
alarm, check the peak value, and change display brightness with the
keys on the front panel (Figure 3).
1. Temporarily
mute the alarm
Figure 3. key functions
Function Descriptions:
1. Temporarily mute the alarm. When the temperature reaches the
alarm temperature, the alarm indicator will flash and meter will start
beeping. Press Shift key once can temporarily mute the alarm. Alarm
will be triggered again if the alarm set temperature is reached again.
Please see Section 6 for detailed explanation alarm settings.
WWW.AUBERINS.COM
3. Switch display
between bright and
dimmed.
2. Check peak value