Auber Instruments SYL-1813-MR Manual de instrucciones

Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida Auber Instruments SYL-1813-MR. Auber Instruments SYL-1813-MR 3 páginas. Multifunction automotive gauge

AUBER INSTRUMENTS
Instruction Manual
A. Specifications
Power supply: 12 V DC (isolated)
Power consumption: < 2 W
Sampling rate: 4 samples/second
Accuracy: 0.2% full scale ±1
Display range: -1999 ~ 9999
Relay contact rating: 3 A @ 220 VAC
Overall dimensions: 60 x 60 x 59 mm
Mounting cutout dimension: 55 mm (2 1/16 inch)
LED display: 0.28'' red color
Working condition: 0 ~ 50 ˚C, ≤ 85% RH
B. Front Panel
Figure 1. Front panel of SYL-1813-MR.
① Display window
② Maximum and Minimum value indicator
③ Alarm and relay J1 indicator
④ Set key
⑤ Shift key
⑥ Up key
Description
1. LED display window to show readings from the sensor.
2. MAX (MIN) on when Display window shows the maximum value or the time of
the Max. MAX (MIN) blinking when Display window shows the minimum value or
the time of Min.
3. AL on indicates alarm is on and J1 relay is pulled in (closed).
4. Set key. To enter parameter setting mode, and to select or save parameters.
5. Shift key (">"). In the parameter setting mode, press this key to select the digit
to be changed. In the normal operation mode, press this key to change the
display in the sequence as shown in the diagram below. Please note that time is
expressed in Hour.Minute (hh.mm) format.
6. Up key ("Λ"). In the parameter setting mode, press it to increase the displayed
value. When displaying Max/Min, press and hold it for 3 seconds to clear the
Max/Min stored. In normal function, this key has two functions, show operation
time and change display brightness. Press and hold it down to show the
operation time since on. Release the key to show the current temperature
(process value). Each time the key is pressed, the display brightness will also be
changed from bright to dim or from dim to bright.
2021.05
SYL-1813-MR Multifunction Automotive Gauge
Version 1.2 (May 2021)
Figure 2. Check the peak values on the gauge by pressing Shift key.
C. Terminal Assignment
Figure 3. Terminal assignment of SYL-1813 gauge.
Description
1. Terminal 1 and 2 are for power input.
2. Terminal 3 is for display brightness control. When connecting the
illumination signal (+12 V) to it, the brightness with synchronized with headlight.
If not connected, the brightness can still be controlled by "Λ" up key.
3. Terminal 4 and 5 are for alarm relay output. See Application example 1
below for how to use it.
4. Terminal 6, 7, 8, and 9 are for different types of input signal. Use 6 and 7 for
thermocouple input. Use 6, 7, and 8 for RTD sensor input. Use 6 and 9 for
pressure sensor.
5. Terminal 10 supplies a regulated +5VDC for powering pressure transducer.
D. Parameter Setting
D.1 Basic parameter (press (set) then input code 0089 to enter)
D.1.A) Basic parameter description
Symbol
Name
Inty
Inty
Dot
Dot
Pul
Pul
PuH
PuH
Psb
Psb
PSbF
PSbF
CorF
CorF
FILT
FILT
End
End
Note 1. These parameters define the scale boundary and resolution of the
display value. They do not apply to temperature sensors.
Note 2. For correcting the offset at zero: Display = measurement + PSb.
This parameter only applies to temperature sensors.
WWW.AUBERINS.COM
Power
12VDC
J1 Output
-
+
5
4
3
2
1
10
9
8
7
6
+ mV mA -
R
+ V -
Regulated
+5VDC
+ TC -
Description
Setting Range
Input Type
See Table 1
Decimal Point
0000 ~ 00.00
Scale Low
-1999 ~ 9999
Scale High
-1999 ~ 9999
Zero offset
-1000 ~ 1000
Range Coefficient
0.500 ~ 2.000
Temperature Unit
C: ˚C; F: ˚F
Digital Filter
0 ~ 3
Exit
Initial
Note
Setting
k
00.00
00.00
1
03.15
0
2
1.000
3
F
0
4
P1/3