3Shape TRIOS S1A-22 Manual de instalación - Página 7

Navegue en línea o descargue pdf Manual de instalación para Equipamiento dental 3Shape TRIOS S1A-22. 3Shape TRIOS S1A-22 13 páginas.

5. Safety and Warnings

5.1 Prerequisites
CAUTION
Read all instructions carefully, including all warnings
and caution notices. You must comply with the warnings
in this manual to prevent injury to persons and damage
to equipment. Proper functionality and safety can only
be guaranteed if the safety precautions in this safety
guide and on the TRIOS IOS system are observed.
Preventive inspection before use of the system
CAUTION
Please examine the TRIOS IOS system for any mechanical
damage on:
• all enclosures,
• all cables.
Safety can only be guaranteed if NO DAMAGE on the TRIOS
IOS system is observed.
Modifications of the system
CAUTION
Modifications of the system which may affect the safety of
the operator, patients or third parties are prohibited by law!
Approved software only
CAUTION
Install only approved software – to prevent interference with
the runtime reliability of the system and programs within it.
Proper training
WARNING
Before you attempt to use the TRIOS scanner equipment
with patients:
• You must be trained to use the TRIOS IOS system and
have read and understood all sections of this guide
describing correct operation.
• You must also be thoroughly familiar with the safe
operation of the TRIOS scanner equipment as described
in this documentation.
In case of equipment failure
WARNING
If at any time the TRIOS IOS system malfunctions,
or if you suspect in any way that the TRIOS IOS system
is not working correctly:
• Remove the scanner from contact with the patient.
• Unplug the system and make sure it cannot be used
before it is checked.
• Contact your reseller.
• Do not attempt to open any covers on the TRIOS IOS system.
Portable RF communications equipment
WARNING
Portable RF communications equipment
(including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches)
to any part of the MOVE/MOVE+ cart system, including
cables specified by the manufacturer. Otherwise, the
performance of this equipment may be impaired.
5.2 Mechanical Hazards
Fragile and movable parts
WARNING
Be careful to avoid the following potential sources of injury:
• Do not lean or sit on any part of the system.
• Movable parts can pinch body parts.
Dropped or damaged equipment
WARNING
If you drop a scanner tip on the floor, you MUST dispose
of it immediately and NOT use the same tip again for
scanning. There is risk that the mirror in the tip has become
dislodged and can fall out.
WARNING
If the handheld scanner is dropped or bumped, it should
immediately be calibrated before further use. If calibration
fails, please contact your technical service provider. See
this guide's instructions on calibrating the handheld scanner.
WARNING
If the handheld scanner is dropped, examine the TRIOS
scanner for any mechanical damage. If a mechanical damage
is observed or suspected, treat this as an equipment failure
(see this guide's instructions on handling equipment failures).
System S1A (MOVE/MOVE+ cart) only
WARNING
• Keep both hands on the upper part of the MOVE/MOVE+
cart when moving it across the floor.
• Do not attempt to lift the MOVE/MOVE+ cart alone.
When moving the system over a door step or other
unsmooth surface, get assistance from another person
to ensure stability.
• When moving the MOVE/MOVE+ cart across the floor,
maintain all its wheels in contact with the floor and
ensure that they are not locked.
• Always place the scanner in its intended mount
on the MOVE/MOVE+ cart when not in use and
when moving the MOVE/MOVE+ cart.
• Keep your feet clear of the MOVE/MOVE+ cart's base
and wheels when moving the MOVE/MOVE+ cart.
• Never drag the MOVE/MOVE+ cart across any of its cables.
• Do not place the MOVE/MOVE+ cart on a slope greater
than 10 degrees.
• Keep sharp objects away from the touch screen.
7