Freedom Innovations Plie 2.0 MPC KNEE Manual de instrucciones - Página 6

Navegue en línea o descargue pdf Manual de instrucciones para Productos de cuidado personal Freedom Innovations Plie 2.0 MPC KNEE. Freedom Innovations Plie 2.0 MPC KNEE 20 páginas. Prosthetic knee

Freedom Innovations Plie 2.0 MPC KNEE Manual de instrucciones

Symbols Legend

The following symbols are found on the Plié® 2.0 MPC Knee:
Attention, refer to Plié® 2.0 MPC Knee Instructions for Use.
Type BF applied part.
Conformité Européene (European conformity).
Federal Communications Commission (USA
Separate collection for electrical and electronic equipment.
The following symbols are found within this Prosthetist Instructions for Use:
Warning! Failure to adhere to this warning may result in user injury, knee damage, or limit the
function of the knee.
Tip
Tip! This information will help the prosthetist and/or user understand the knee.

Safety Information

Prior to use, the prosthetist should discuss the following warnings with the user to prevent injury.
Without power, the Plié® 2.0 MPC Knee will not release into swing phase. In this powered-down mode, the
knee will default to the stance flexion resistance setting. Insert a charged battery to resume normal knee
function.
Use only the batteries and battery charger intended for use with the Plié® 2.0 MPC Knee.
Extreme temperatures will affect the performance of the Plié® 2.0 MPC Knee. Allow sufficient time for the
knee to return to temperatures within the specified operating temperature range before use.
High elevations and cold temperatures will affect the knee internal air pressure. In these conditions, the user
may be required to increase the air pressure to resume normal function.
The Plié® 2.0 MPC Knee requires air pressure for normal operation. A cylinder air pressure of less than 20
PSI may detrimentally affect the stance flexion resistance. The user should be provided with the included air
pump and instructed of its proper use. The user will be required to maintain sufficient air pressure by
periodically recharging the air reservoir, using the provided air pump. The cylinder air pressure should not
exceed 100 PSI.
The threaded cap for the swing flexion resistance should remain threaded in the port except when the user is
adjusting the air pressure. The threaded cap functions as a secondary seal and prevents lint and debris from
contaminated the air valve seat. The threaded cap should be gently tightened with fingers or a 4mm allen key.
If the user changes shoe heel height, the alignment may become unstable. An adjustable heel height
prosthetic foot is recommended or the user should be provided with heel wedges to accommodate various
heel heights.
If the user changes shoe heel height, the Plié Control Threshold Setting may require adjustment as the knee
requires a minimum user toe loading (adjustable Toe Threshold) to release into swing phase.
The Plié® 2.0 MPC Knee requires calibration of the load and angle sensors. Loss of sensor calibration may
prohibit the knee from normal function. Re-calibrate the sensors to resume normal function.
Plié® 2.0 MPC Knee Instructions for Use (R-720-109 Rev. D), English!
conformity).
!
!
!
!
6