 Isopropanol (70%)
 Diluted ammonia water
 Hydrogen peroxide (3%)
 Alcohol
 Ethanol (70%)
 Diluted formaldehyde (35%-37%)
 Diluted sodium hypochlorite solution (bleaching agent)
NOTE
Sodium hypochlorite solution with a concentration of 500ppm (1:100 diluted bleacher
solution used in family) ~ 5000ppm (1:10 diluted bleacher solution used in family) is
very effective. How much ppm depends on how much organic matter (blood,
propagation grime etc.) existing on the surface.
CAUTION
 Never use strong solvent, such as acetone.
 ALWAYS dilute the solutions according to the manufacturer's suggestions.
 NEVER use abrasive, erosive cleaners, or cleaners containing acetone.
 NEVER permit fluids run into the casing, switches, or openings between two cushions
in the device.
 NEVER submerge the device into water or any cleaning solution, or pour or spray
water or any cleaning solution on the device.
 ALWAYS wipe off all the cleaning solution with a dry cloth after cleaning and dry the
device in the air. Never dry the device in the violent sunshine or toast it under high
temperature.
 If the device is polluted by chemical substance, the users should handle it effectively
according to the properties of the chemical substance.
WARNING
The cleaning solutions above can only be used for general cleaning. If you use them to
control infections, the manufacturer shall assume no responsible for the effectiveness.
Please consult your hospital's infection controllers or professionals.
Disinfection
Disinfection may cause damage to the device. We recommend the disinfection is contained
in the hospital's servicing schedule only when necessary. The device should be cleaned prior
to disinfection.
Recommended disinfection material: Alcohol based (Ethanol 70%, Isopropanol 70%), and
aldehyde based.
CAUTION
 ALWAYS dilute the solutions according to the manufacturer's suggestions and adopt
lower concentration if possible.
 NEVER submerge the device into water or any solution, or pour water or any solution
on the device.
 ALWAYS wipe off all the excess liquids on the device surface with a dry cloth.
 Never use EEO and formaldehyde to disinfect.
 Never permit high-pressure and high-temperature disinfection of the device.
WARNING
Disinfection may cause damage to the device; therefore, when preparing to disinfect
the device, consult your hospital's infection controllers or professionals.
11