Apogee SP-214 Manual del usuario - Página 7
Navegue en línea o descargue pdf Manual del usuario para Accesorios Apogee SP-214. Apogee SP-214 19 páginas. Pyranometer
También para Apogee SP-214: Manual del usuario (18 páginas)
Spectral Response
Temperature Response
Spectral response estimate of Apogee
silicon-cell pyranometers. Spectral
response was estimated by multiplying
the spectral response of the
photodiode, diffuser, and adhesive.
Spectral response measurements of
diffuser and adhesive were made with a
spectrometer, and spectral response
data for the photodiode were obtained
from the manufacturer.
Mean temperature response of four
Apogee silicon-cell pyranometers.
Temperature response measurements
were made at approximately 10 C
intervals across a temperature range of
approximately -10 to 50 C under
sunlight. Each pyranometer had an
internal thermistor to measure
temperature. At each temperature set
point, a reference blackbody
pyranometer was used to measure solar
intensity.