Sensor Calibration
Apogee amplified quantum sensor models have a standard PPFD calibration factor of exactly:
Multiply this calibration factor by the measured mV signal to convert sensor output to PPFD in units of µmol m
Calibration Factor (0.5 µmol m
0.5
Full Sunlight
(2000 µmol m
-2
s
-1
)
SQ-212, SQ-222: 1.0 µmol m
SQ-215, SQ-225: 0.5 µmol m
s
per mV) * Sensor Output Signal (mV) = PPFD (µmol m
-2
-1
*
Example of PPFD measurement with an Apogee quantum sensor. Full sunlight yields a
PPFD on a horizontal plane at the Earth's surface of approximately 2000 µmol m
yields an output signal of 4000 mV for the 0-5 V option or an output signal of 2000 mV for
the 0-2.5 V option. The signal is converted to PPFD by multiplying by the calibration
factor.