Apogee SI-421 Manual del usuario - Página 11
Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida Apogee SI-421. Apogee SI-421 20 páginas. Infrared radiometer
También para Apogee SI-421: Manual del usuario (20 páginas), Manual del usuario (20 páginas)
11
Apogee SI series infrared radiometers are calibrated in a temperature controlled chamber that houses a
custom-built blackbody cone (target) for the radiation source. During calibration, infrared radiometers
(detectors) are held in a fixture at the opening of the blackbody cone, but are thermally insulated from
the cone. Detector and target temperature are controlled independently. At each calibration set point,
detectors are held at a constant temperature while the blackbody cone is controlled at temperatures
below (12 C), above (18 C), and equal to the detector temperature. The range of detector temperatures is
-15 C to 45 C (set points at 10 C increments). Data are collected at each detector temperature set point,
after detectors and target reach constant temperatures.