Apogee Instruments SIF-1H1 Manual del usuario - Página 7

Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida Apogee Instruments SIF-1H1. Apogee Instruments SIF-1H1 19 páginas. Infrared radiometer
También para Apogee Instruments SIF-1H1: Manual del usuario (17 páginas)

Apogee Instruments SIF-1H1 Manual del usuario
Spectral Response
Spectral response of SIF series
infrared radiometers. Spectral
response (green line) is determined
by the germanium filter and
corresponds closely to the
atmospheric window of 8 to 14 µm,
minimizing interference from
atmospheric absorption/emission
bands (blue line) below 8 µm and
above 14 µm. Typical terrestrial
surfaces have temperatures that
yield maximum radiation emission
within the atmospheric window, as
shown by the blackbody curve for a
radiator at 28 C (red line).