6 Troubleshooting
When the remedy for some problem cannot be discovered easily, the
recording of communication between Clarity and the autosampler can
significantly help the DataApex support to discover the cause of the
problem.
The recording can be enabled by adding or amending the COMMDRV.INI
file in the Clarity installation directory (C:\CLARITY\CFG by default). The file
can be edited in any text editor (e.g. Notepad). Following section should be
edited or added:
[COM1]
echo=on
textmode=on
filename=SparkMarathon_%D.txt
reset=off
Note:
Instead of COM1 type the correct serial port used to communicate with
the Spark Marathon autosampler. This port number is displayed when
the AS Status button in the
Spark Marathon Setup
Note:
%D (or %d) in the filename parameter means that the log will be created
separately for each day. The reset=off parameter disables deleting the
content of the log each time the station is started during the same day.
The created *.TXT files will greatly help in diagnosis of unrecognized errors
and problems in communication.
Method Setup - AS
dialog.
- 13 -
6 Troubleshooting
dialog is invoked or in the