6 Troubleshooting

When the remedy for some problem cannot be discovered easily, the
recording of communication between Clarity and the detector can
significantly help the DataApex support to discover the cause of the
problem.
The recording can be enabled by adding or amending the COMMDRV.INI
file in the Clarity installation directory (C:\CLARITY\CFG by default). The
file can be edited in any text editor (e.g. Notepad). Following section
should be edited or added:
[TCP_IP 192.168.0.100:79]
echo=on
textmode=on
filename=Alltech3300ELSD_%D.txt
reset=off
Note:
Instead of 192.168.0.100:79 type correct IP address and port used for
communication with the Alltech 3300 ELSD detector. This port number
is displayed when the Det Status button in the
dialog is invoked.
Note:
%D (or %d) in the filename parameter means that the log will be created
separately for each day. The reset=off parameter disables deleting the
content of the log each time the station is started during the same day.
The created *.TXT files will greatly help in diagnosis of unrecognized
errors and problems in communication.