4.3.
Setup.
If the moving table is used, unpack it carefully and assemble according to the
-
enclosed assembling scheme.
Place the unit upon stable surface. Ensure that the unit is placed on a solid, level
-
surface not less than 720x550 mm (1290x600 mm for UVT-S-AR model), which is
able to support its weight and the weight of equipment and materials inside, for
instance on the T-4 / T-4L table.
Connect the power cable to the socket on the rear side of the unit, position it with
-
easy access to the power switch and plug.
4.4.
Shelf setup. Refer to the instructions included with the shelf.
5.
Operation
Note. For UVC/T-M-AR and UVT-B-AR boxes, check that the short power
cable (fig. 5/1) is connecting the fuse box and the controller top.
5.1.
Connect the power plug to a grounded power socket. Turn on the Power switch:
UVC/T-AR – no switch, the unit powers up automatically;
UVC/T-M-AR, UVT-B-AR – on the right side of the unit, behind the rear wall (fig. 5/3);
UVT-S-AR – on the left side of the unit, behind the rear wall (fig. 6/1).
5.2.
UV exposition of the working place.
Caution! The front protective screen must be closed while the open UV lamp is
5.2.1. Turn ON the switch 2 (fig. 2/2) on the control panel (fig. 3/1). This switches on the UV
recirculator inside the unit (fig. 3/2) and the open UV-lamp exposition timer. The UV
recirculator will operate all the time until the switch 2 is turned OFF.
8
switched on. Otherwise, the operator can be exposed to a dangerous level
of UV emission.
Figure 2. Control panel