CMP 6050 RETROFIT INSTALLATION INSTRUCTIONS
16.
F
Following schematics provided for optional chamber features (257349), make connections for all
existing chamber options to the control panel. Pay careful attention to wires shown to be removed
from original wiring (original wiring indicated by dashed-line boxes).
F
Electrostatic
Discharge
WARNING
Existing driver boards
•
Both styles may be
present and must be
removed
Existing driver boards
•
Before removal
Driver board cable
•
With attached female
connector
18.
F
Using original chamber schematics provided, locate existing wire labelled 1 leaving the power
supply breaker and follow it to the 1 terminal at terminal block 3 (TB3) (Figure 2). Remove it from
the 1 terminal and wire to the 3 terminal and remove wires to the start/stop switch. If use of the
start/stop switch is required, run the wire labelled 1 from the power supply breaker to the start/stop
switch replacing the wire labelled 2, and remove the wire labelled 2 from the panel.
6
May 2013 |
257775R00
17.
Electrostatic Discharge WARNING: Wear included electrostatic wrist guard
and observe procedures for handling static sensitive equipment.
Remove all existing control driver boards from output boards (Figure 12 and Figure
13). Remove new control driver boards from static sensitive packaging and connect
to control driver board cables hanging from lower wire bundle using attached black
plastic connectors (Figure 14). Using labels as a guide, insert all new control driver
boards into their respective slots. Use caution not to bend delicate contact pins
when inserting driver boards.
Figure 12
Figure 13
Figure 14
Communication ports
•
Communications cables
removed in Step 6.
New control driver
board
•
With attached male
connector
CONVIRON