COOK Medical Duette DT-6-5F Manual de consulta rápida
Navegue en línea o descargue pdf Manual de consulta rápida para Equipamiento médico COOK Medical Duette DT-6-5F. COOK Medical Duette DT-6-5F 2 páginas. Multi-band mucosectomy
Duette
®
M U LT I - B A N D M U C O S E C T O M Y
Scope Compatibility
Endoscope Outer Diameter
Order No.
DT-6-5F
9.5 mm - 13 mm
DT-6
DT-6-XL
11 mm - 14 mm
Prep 3
Introduce either end of the loading catheter
through the white seal in the ligator handle and
advance, in short increments, until it exits the tip
of the endoscope.
M E D I C A L
cookmedical.com
Prep 1
The firing position allows the handle to be
rotated in the forward direction only. The
two-way position allows the handle to rotate
in both directions. Prior to introducing the
endoscope, keep the handle in the two-way
position.
Prep 4
Attach the trigger cord, leaving approximately
2 cm of cord between the knot and the hook.
Withdraw the loading catheter and trigger cord
up through the endoscope and out through the
ligator handle.
Q U I C K R E F E R E N C E G U I D E
Prep 2
Olympus
Pentax
Insert the ligator handle into the endoscope
accessory channel.
Note: The irrigation adapter may be inserted
through the white self-sealing valve prior to
introducing the loading catheter.
Prep 5
Attach the barrel to the tip of the endoscope,
advance as far as possible.
Fujinon