Maintenance
The ACCU-CAL™ 50 was designed to operate with minimum
maintenance. Follow the schedule below to assure top unit
performance.
•
Calibrate the instrument at least annually. Calibration
service is available through Dymax Customer Service or
Dymax Product Repair.
•
Change the batteries when a low battery warning is
received. The ACCU-CAL™ 50 uses two AA-type
batteries. The battery compartment is on the back of the
instrument.
•
Keep the detector head's sensing element clean and free
of contaminants. The detector head may be cleaned with
a clean tissue wetted with isopropyl alcohol.
•
To increase battery life, turn unit power off after use. The
unit will not automatically power off.
Spare Parts and Accessories
Item
Adapter Kit – Flood to Spot Model (includes parts listed below)
Lightguide Simulator
8-mm Lightguide Adapter
5-mm Lightguide Adapter
3-mm Lightguide Adapter
Specifications
Property
Power
Display
Detector Interface
Measurement
Ranges
Front Panel Control
Temperature
Size
Weight
8
Specification
Two AA size batteries
LCD Graphic Display 97x 32 Pixel
Display area: 0.56 in x 1.41 in (14.3 mm x 35.8 mm)
9-Pin MDSM9 socket, 4 inputs
Four modes of operation
Auto range within each operating mode
3 buttons
Operating: 5 to 40°C
Storage: -10 to 50° C
5.71 in x 2.48 in x 1.18 in (145 mm x 63 mm x 30 mm)
0.33 lbs (150 g)
Figure 7.
Battery Compartment (Closed & Open)
Part Number
U S E R G U I D E
A C C U - C A L ™ 5 0 R A D I O M E T E R
|
39554
38408
39558
39557
39556