Tubing Installation

Figure 3.
Tubing Installation Procedure
B - Tube
Support
Note: The Model 828 is not supplied with tubing. Various tubing sizes and materials
(including polyethylene, polyurethane, PVC, and silicone) are available for order
through Dymax. This valve is compatible with a maximum tubing size of 0.375" [9.53
mm] OD.
1.
Separate the Tube Support (Figure 3
removing the four Socket Head Cap Screws (Figure 3
2.
Place the Tubing (Figure 3, C) in the slot (Figure 3, E) on the Tube Support.
3.
Reconnect the Tube Support to the Valve Body by reinstalling the four Socket
Head Cap Screws.
9
A - Valve Body
E - Slot
D - Socket Head Cap
Screws
,
A) from the Valve Body (Figure 3, B) by
,
U S E R G U I D E
C - Tubing
D).
|