Chapter 9
Specification
Test Sample
Fresh Capillary Whole Blood
Measuring Time
GLU:5 seconds; CHOL: 26 seconds; UA: 15 seconds
Measuring Range
GLU: 20-600 mg/dL (1.1 - 33.3 mmol/L);
CHOL: 100 - 400 mg/dL ( 2.59 - 10.35 mmol/L );
UA :3 - 20 mg/dL ( 0.18 - 1.19 mmol/L )
Sample Volume
GLU: 0.7 μL; CHOL: 0.8 μL; UA: 1 μL
Storage & Transportation
4 - 30˚C (39-86˚F)
Condition
Operation Temperature
10 - 40˚C (50-104˚F)
Storage & Transportation
10-90%
Relative Humidity
Memory
460 Test Results (GLU: 360 ; CHOL: 50 ; UA: 50)
Battery Type
One 3V (CR2032) lithium battery
Battery Life
Approximately 1,000 tests
Dimensions
88*52*16 mm
Weight
About 50g (with battery)
Altitude
10,000 feet (3048 m)
Expected Service Life
5 years
Transmission Function
Bluetooth 4.0
Note:
• Please refer to the strip insert for accuracy, precision, limitation, and other important   
 information.
25
Electromagnetic Compatibility
This meter meets the electrostatic discharge immunity testing was basic standard
IEC6100-4-2. In addition, the meter meets the electromagnetic emissions requirements as
per EN61326. The purpose of these requirements is to ensure that meter does not affect or
being affected by other devices during operation.
Warning :
● Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally. The BeneCheck Multi-Monitoring Meter may provide a
description or list of equipment with which the BeneCheck Multi-Monitoring
Meter has been tested in a stacked or adjacent configuration.
● Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
● Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the BeneCheck Multi-Monitoring Meter, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.