Active Motif PIXUL Manual del usuario - Página 9

Navegue en línea o descargue pdf Manual del usuario para Equipamiento médico Active Motif PIXUL. Active Motif PIXUL 15 páginas. Multi-sample sonicator

Installing the PIXUL Multi-Sample Sonicator

The following instructions detail the installation procedure for the PIXUL Multi-Sample Sonicator.
Only attempt to install this instrument after reading these instructions.
No special tools are required for installation.
The PIXUL instrument is intended for indoor use, and the power plug must be grounded.
Place the PIXUL instrument on a level, non-slip surface with sufficient load bearing capacity. Do
not place anything heavy on the PIXUL instrument.
Do not cover the ventilation openings. Provide at least 20 cm of clearance to allow proper airflow.
Toxic vapors can be produced depending on the type of operation. Ensure sufficient extraction of
the vapors.
1. Open the boxes and cut the shipping straps. Remove the molded styrofoam packaging, protec-
tive clear wrapping, and any tape securing moving parts. Inspect the instrument and accessories
immediately after delivery for completeness and transport damage.
2. Place the PIXUL instrument on a bench. Always make sure the PIXUL is on a level surface.
3. Fill the reservoir of the PIXUL instrument with approximately 0.5 liters of PIXUL Coupling Fluid.
The fluid level should reach about an inch below the reservoir top. If you accidentally overfill the
reservior, wipe up any spilled Coupling Fluid and suction Coupling Fluid out of reservoir until it
reaches the appropriate level.
4. Plug the power cord into the back side of the PIXUL instrument and the wall outlet.

Operating the PIXUL Multi-Sample Sonicator

www.activemotif.com
6