DAYCO D165DC Manual del usuario - Página 10

Navegue en línea o descargue pdf Manual del usuario para Herramienta de engarce DAYCO D165DC. DAYCO D165DC 19 páginas.

CALIBRATION PROCEDURE
When the crimper is correctly calibrated, the ram will extend and fully
close the die set. After the die set is fully closed, the time from which
the pump starts to build pressure and the point at which the motor
shuts of automatically will be approximately 1 second.
Many problems associated with incorrect crimp diameters are caused
by incorrect calibration.

CALIBRATION

Insert the Standard Pressure Plate, any die
set, and the compression ring in the order
illustrated.
Set the Micro-Crimp adjuster at "95"
Press and hold the start Switch
If the ram extends closing the dies to their fully closed position
and the motor shuts off approximately 1 second after the pump starts
to build pressure (the sound of the pump will change), then the crimper
is correctly calibrated.
If the crimper requires re-calibration, hold the
micrometer barrel with a 5/16 inch open end wrench and
rotate the stem either in or out with a 5/32 inch hex key.
If the time from which the pump starts to build pressure is
greater than approximately 1 second, rotate the stem out
slightly.
If the time is less than approximately 1 second, rotate the
stem in slightly.
Recheck calibration.
Page
10