MEASURING CABLE REPLACEMENT
1.
If measurement cable is still intact, cut cable end off and discard.
2.
Mark the end plate pot side (side with electrical wires) relative to the housing with
a pen/pencil. A simple line starting on the end bracket and continuing to the edge
of the housing will do.
3.
Loosen the set screws on the end plate pot side using a 0.050" Allen wrench.
4.
Gently remove the end plate pot side and set aside. Be careful not to lose the tiny
ball bearings, which will either be on the spool shaft or in the center of the end
plate.
The spool is now loose. If allowed to, it will fall out and the spring will pop out.
Hold the spool to the side plate by pinching the spool and side plate together
through step 7. If the spring comes out of the housing use your fingers to rewind
the spring back into the housing, being careful not to kink the spring.
5.
Loosen the set screws holding the side plate using a 0.050" Allen wrench.
6.
Push the spool and side plate out of the housing while holding together.
7.
Insert a small flathead screwdriver between the side plate and the spool and gently
slide the spring off the spring arbor (the spring arbor is part of the spool). Slide
from the spool toward the side plate. Using the screw driver to keep the spring in
place in the side plate, separate the spring and side plate. Be careful not to lose
the tiny ball bearings either on the spool shaft or in the hollow in the side plate.
8.
Remove and discard the cable crimp, which is glued to the spool. It will have be
pried off using a screwdriver or other sharp object.
9.
Discard any remaining cable.
10.
Scrape off any remaining glue left on the spool.
11.
Cut new cable kit to length. For a 3 inch MTA the cable should be 3.5 +/- .125.
For the 5 inch MTA the cable should be 5.5 +/- .125. This length is measured
from the bottom edge of the snubber, (see diagram).
12.
String cable through the cable hole in the housing and then through the cable hole
in the spool.
13.
Once the cable is through BOTH holes fasten the crimp to the cable. Line the
crimp up with the end of the cable and crimp completely flat using the flat portion
Page 2 of 4
IMPORTANT