processing section is communicating with the internal host
processor.
5. From the Home page, press Preferences > Display Preferences.
From the Display Setup Menu, select the Calibrate Touch Screen
button. The user will be prompted to follow the instructions. Once the
touch screen is successfully cablibrated, press OK and then Accept
the new calibration settings. When calibration is complete, Exit the
setup screens until the Home page appears.
NOTE: Use a stylus or a blunt object like a pencil eraser when
working on the touch screen. Do not use a pointed object as it may
damage the screen.
6. From the Home page, press Editor > Assign a 5D to a slot, to check
if the 5D module installed matches the required physical placement
of modules at the rear panel of the 4500, as noted in Step 1. If there
is a conflict or to double-check 5D module placement, press the Get
12 |
Daytronic Model 4500 Quick Reference Guide