5.6. Battery autonomy estimations
Measuring channels
3D
3D Gyroscope
Accelerometer
256Hz
128Hz
128Hz
256Hz
256Hz
512Hz
512Hz
*Autonomy measured at room temperature
5.7. Troubleshoot
Soft reset
The soft reset is the manipulation which can be done if the Physilog® is not working properly. To do
a soft reset, hold the main button for 15 seconds. The LED shines green, turns off and finally blinks
red/orange and then stops. If the problem persists, fill out the contact form on the support page
(www.gaitup.com/support) or write an email to: [email protected]. Please indicate Physilog®
serial number and describe the problem in as much detail as possible (number of red blinks, last
manipulation, firmware version etc.).
Hard reset
If the soft reset doesn't allow to recover Physilog®, a hard reset is necessary. A hard reset is done
when the sensor's battery goes completely out (few days to several weeks depending on firmware
version and charging state). To force a hard reset, introduce a thin item (like the end of a paperclip)
into the hole besides the USB entry and gently press until you feel a button pressed inside the sensor.
Warning: This manipulation destroys the waterproofing membrane and the sensor is therefore no
Physilog5 user manual v1.2.6 Date: 19.06.2018
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
Barometric
pressure
256Hz
64Hz
128Hz
64Hz
128Hz
64Hz
256Hz
64Hz
256Hz
64Hz
512Hz
64Hz
512Hz
64Hz
EPFL Innovation Park, Bat C. CH-1015 Lausanne
Radio settings
Firmware
Master / slave /
Version (release
no radio
-
(02.03.2017)
Master / slave
pairs / no radio
(14.06.2017)
Slave without
master
(14.06.2017)
Master / slave
pairs / no radio
(14.06.2017)
Slave without
master
(14.06.2017)
Master / slave
pairs / no radio
(14.06.2017)
Slave without
master
(14.06.2017)
www.gaitup.com
TVA CHE-145.136.760
Approx.
version
Autonomy *
Hours
date)
1.0.2
18h
1.1.0
20h
1.1.0
10h
1.1.0
18h
1.1.0
10h
1.1.0
15h
1.1.0
10h
Gait Up SA