Lepu Medical LFR30B Manual de instalación y uso - Página 5

Navegue en línea o descargue pdf Manual de instalación y uso para Termómetro Lepu Medical LFR30B. Lepu Medical LFR30B 8 páginas.

Lepu Medical LFR30B Manual de instalación y uso
Shenzhen LEPU Intelligent Medical Equipment Co., Ltd.
The infrared forehead thermometer has been calibrated before delivery from the factory. If
you have any doubt about its accuracy, please contact the after-sales service.
We recommend technical inspection for the measurement every two years, and it is required
to comply with the applicable national regulations of the local place. Technical inspection for the
measurement may be carried out by the government agencies in charge or by authorized
fee-paying maintenance services.
1.16 Measured temperature too high
Error message on screen
1.17 Measured temperature too low
Error message on screen
1.18 Ambient temperature out of range
Error message on screen
Form No.: LP-QR-RD-016 REV.01

Troubleshooting

Meaning of show
Measured body temperature is
too high (above 43.0°C)
Measured object temperature is
too high (above 100°C)
Meaning of show
Measured body temperature is
too low (below 32.0°C)
Measured object temperature is
too low (below 0°C)
Meaning of show
Possible causes and solutions
The result exceeds the
measurement range limit
The result exceeds the
measurement range limit
Possible causes and solutions
The result exceeds the
measurement range limit
The result exceeds the
measurement range limit
Possible causes and solutions
Page 5of 7