iHealth BP3 Manual del usuario - Página 10
Navegue en línea o descargue pdf Manual del usuario para Tensiómetro iHealth BP3. iHealth BP3 12 páginas. Fully automatic arm cuff electronic blood pressure dock
También para iHealth BP3: Manual de instrucciones (11 páginas), Manual de inicio rápido (2 páginas)
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacture's declaration - electromagnetic emissions
The BP3L is intended for use in the electromagnetic environment specified below.
The customer or the user of the BP3L should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The BP3L is intended for use in the electromagnetic environment specified below. The customer or the user of the BP3L
should assure that it is used in such an environment.
IMMUNITY test
IEC 60601test level
Electrostatic discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply
transient/burst
lines
IEC 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
<5 % U
interruptions and
(>95 % dip in U
voltage variations
for 0.5 cycle
on power supply
40 % U
input lines
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Compliance
Electromagnetic environment - guidance
The BP3L uses RF energy only for its internal function. Therefore, its RF
Group 1
emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The BP3L is suitable for use in all establishments, including domestic
Class B
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
Class A
purposes.
Complies
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5 % U
T
)
(>95 % dip in U
T
for 0.5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
T
)
typical commercial or hospital environment. If
T
the user of the BP3L requires continued
operation during power mains interruptions, it
T
is recommended that the BP3L be powered
)
T
from an uninterruptible power supply or a
battery.
T
)
T