iHealth BP5 Manual del usuario - Página 8
Navegue en línea o descargue pdf Manual del usuario para Tensiómetro iHealth BP5. iHealth BP5 17 páginas. Wireless blood pressure monitor
También para iHealth BP5: Manual de inicio rápido (4 páginas), Manual del usuario (13 páginas), Manual de inicio rápido (5 páginas)
device be set in Airplane mode during measurement to avoid strong
magnetism interference.
6. If Irregular Heartbeat (IHB) is detected during the measurement
procedure, the IHB symbol will be displayed. Under this condition,
the Wireless Blood Pressure Monitor can keep functioning, but the
results may be inaccurate. Please consult your physician for accurate
assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s and the number of
such pulse takes more than 53 percent of the total number of pulses.
7. Please do not use any other cuff other than that supplied by the
manufacturer as this may result in measurement errors and a
biocompatible hazard.
8. Information regarding potential electromagnetic or other
interference between the blood pressure monitor and other devices
together with advice regarding avoidance of such interference,
please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is
suggested that the blood pressure monitor be kept 10 meters away from
other wireless devices, such as WLAN unit, microwave oven, etc.
9. This product should not be used as a USB device.
10.This product is verified by auscultatory method. It is recommended
that you check Annex B of ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for
verification method details if needed.
11.If the determined blood pressure (systolic or diastolic) is outside
the rated range specified in part SPECIFICATIONS, the app will
immediately display a technical alarm on screen. In this case, consult
a physician or ensure that proper measurement procedures are followed.
The technical alarm is preset in the factory and cannot be adjusted
or inactivated. This technical alarm is assigned as low priority
according to IEC 60601-1-8.The technical alarm is non-latching and does
not need to be reset.
12.A medical AC adapter with an output of DC 5.0V and complies with
IEC 60601-1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2 is suitable for
this monitor, such as ASP5-05010002JU (input: 100-240V, 50/60Hz, 200mA;
output: DC 5.0V, 1.0A). Please note that the monitor jack size is USB
mini B.
This Wireless Blood Pressure Monitor is designed for adults and
should never be used on infants, young children, pregnant or
pre-eclamptic patients. Consult your physician before use on
children.
This product might not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
Please do not share the cuff with any infectious person to avoid
cross-infection.
8