professional.The technical alarm is preset in the factory and cannot
be adjusted or inactivated. This technical alarm is assigned as low
priority according to IEC 60601-1-8. The technical alarm does not
need to be reset.
This Wireless Blood Pressure Monitor is designed for adults and
should never be used on infants, young children, pregnant or
pre-eclamptic patients. Consult your physician before use on
Children.
This product might not meet its performance specifications if stored
or used outside the specified temperature and humidity ranges.
Please do not share the cuff with any infectious person to avoid
cross-infection.

TROUBLESHOOTING

PROBLEM
LCD shows low battery symbol
LCD shows "Er 0" or
APP display reads "ERROR"
LCD shows "Er 1"or
APP display reads "ERROR"
LCD shows "Er 2"or
APP display reads "ERROR"
LCD shows "Er 3"or
APP display reads "ERROR"
LCD shows "Er 4"or
APP display reads "ERROR"
LCD shows "Er 5""or
APP display reads "ERROR"
LCD shows "Er 6""or
APP display reads "ERROR"
LCD shows "Er 7""or
APP display reads "ERROR"
LCD shows "Er 8""or
APP display reads "ERROR"
LCD shows "Er A""or
APP display reads "ERROR"
Bluetooth connection unstable
POSSIBLE CAUSE
Low Battery
Pressure system is unstable
before measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cuff
is too tight during inflation
Pneumatic system leakage or
cuff is too loose during inflation
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter error
Bluetooth
connection
unsuccessful,
monitor
abnormal,
or
electromagnetic interference is
present
第 9 页 共 16 页
SOLUTION
Change the battery
Don't move and try again.
Apply the cuff correctly and
try again
Measure again after five
minutes. If the monitor is
still abnormal, please
contact the local distributor
or the factory.
Reset iOS/Android device
and take out batteries for
five minutes, and then
reinstall all batteries.. Make
is
sure the monitor and
strong
iOS/Android device are
away from other electrical
equipment. Please see
GENERAL SAFETY