iHealth Neo BP5S Manual del usuario - Página 14
Navegue en línea o descargue pdf Manual del usuario para Tensiómetro iHealth Neo BP5S. iHealth Neo BP5S 14 páginas.
810
800-960
870
930
1720
1700-1990
1845
1970
2450
2400-2570
5240
5100-5800
5500
5785
Phenomenon
Electrical
fast
transients/burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted
disturbances induced
by RF fields
Voltage dips
Voltage interruptions
Date of issue:
Jul
4, 2017
BP5S-SMSY01 V1.0
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 9V/m
Table 4 – Input a.c. power Port
Basic EMC standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
14 / 14
Immunity test levels
Home Healthcare Environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV, ±2 kV
3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands between 0.15MHz
and 80MHz
80%AM at 1kHz
0% U
; 0.5 cycle
T
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% U
; 1 cycle
T
and
70% U
; 25/30 cycles
T
Single phase: at 0º
0% U
250/300 cycles
T;