At the same time, the Inhale Valve (G) is opened by the expiratory pressure created by releasing the bag,
and send the air into the bag through the top of the Reservoir Valve, and at the same time, send the
Oxygen into the bag from the Oxygen Reservoir till the bag returns to the original shape before
compressing. To avoid excessive Oxygen flow rate and low compressing frequency causing too high
pressure inside the bag and the Reservoir, the Reservoir Valve (I) is specially designed to release the
excessive air, to keep a low‐rate Oxygen supply and ensure the patient's safety
Principle Drawing (D)
GaleMed reusable resuscitators are equipped with pressure relief valves (optional), automatically provide
and adjust the pressure in the lung, and keep it within 40 / 60 cmH
will cause the pressure relief valve to jump off and push the pressure out to ensure the patient's safety.
Should higher inspiratory pressure be required the pressure relief valve may be overridden by placing the
thumb ever the valve as show as follows.
3.1.8 Direction for Use:
3.1.8.1 Reusable‐Resuscitator Operation
1. Place the patient on back, pull his chin upward as
possible to keep the airway and the mouth cavity in
alliance line, so the patient can breathe smoothly.
2. Clean all visible foreign material inside the mouth and
the throat.
3. Insert the oropharyngeal tube, keep the patient's mouth
open to prevent tongue from occluding the airway. (Can
use a mouth opener to open his mouth) The
oropharyngeal tube can be selected according to the
patient's mouth cavity size.
4. The emergency personnel should stay behind the
patient's head, extend the head back and pull his chin
upwards and towards the emergency personnel.
Pressure
Relief Valve
Pressure Relief Valve Motion
Pressure
Relief Valve
Overriding the Pressure Relief Valve
O, any pressure exceed this standard