Afinion AS100 Manual rápido - Página 2

Navegue en línea o descargue pdf Manual rápido para Instrumentos de medida Afinion AS100. Afinion AS100 4 páginas.

Afinion AS100 Manual rápido
Afinion
ACR
TM
Prepare for Testing
2
Switch the Analyzer on.
Allow 15 minutes for the Test Cartridge to reach operating temperature
20-30°C (64-86°F) before use.
Open the pouch just before use. Hold the Test Cartridge by the handle.
Label the Test Cartridge with sample ID. Use the ID area.
Analyze Afinion™ ACR Control before analyzing patient samples.
Procedure for Collecting the Sample
3
Sampling from a Control Vial
Follow the procedure described below. See page 4 for control testing recommendations.
Sampling from a Urine Cup
Always use gloves.
Patient samples stored refrigerated can be used without equilibration to room temperature.
Mix the sample material well by inverting the urine cup 8–10 times before collecting the sample.
The sample can be extracted from the urine cup or the cap.
Specimen Collection using the Afinion™ ACR Test Cartridge
1
Pull up the sampling device.
4
(a)
(b)
Avoid air bubbles and
incomplete filling (a). Avoid
sample on the outside of
the capillary. Do not dip the
neck of the sampling device
into the urine cup (b). Do not
wipe off.
2
(a)
(b)
Touch the surface of the patient
sample
(a)
or control (b).
5
Insert the sampling device
immediately.
Quick Guide
Tear strip
Handle
3
Fill the capillary end to end.
It is not possible to overfill.
6
Within 1 minute place the
Test Cartridge in the Analyzer.
2