Jadever JWL-1.5K Manual - Página 10
Navegue en línea o descargue pdf Manual para Accesorios Jadever JWL-1.5K. Jadever JWL-1.5K 20 páginas.
También para Jadever JWL-1.5K: Manual (20 páginas)

RTC setting:
When the window shows "on", press
/MR
year. Press
/MR
Press
or
HOLD/LO/-
to change the value. Press
17. P 16:
display the previous weight when powering on again
ON
OFF not display the previous weight when powering on again
18. P17:
M-/SET
Press
6.Calibration
Note: Before calibration, please set the unit first. The unit used in calibration is the one that has
been set before. During the calibration procedure, press
mode without saving.
Here we take 3kg/1g as an example
1. Press and hold
2. With no load on the weighing pan, press
blanking at the left bottom.
3. Wait till the window displays the first calibration value. "
Note: The first calibration value is default. With the same capacity, the last first calibration point
value can be recorded. If the capacity has been changed, the default value is 1/3 of full load. If
you need to change the value, do as the following: Press
/MR
Press
HOLD/LO/-
to change the value. Press
TARE/
or
to choose date and time. Press
TARE/
to shift key leftward or rightward; Press
Zero-Offset function.
Initialization
M+/SAVE
then press
TARE
while powering on. Do not release it till the window displays "CAL".
TARE/
or
to move leftwards or rightwards. Press
(II)
M-/SET
to enter RTC setting and the window shows the
M+/SAVE
to return.
to initialization and the window display RESET.
ZERO/ESC
TARE
to start zero point calibration.
M+/SAVE
to save.
JWL
M-/SET
to enter setting.
+/PRINT/HI
to return to normal weighing
"appears at the left bottom.
M-/SET
to enter the value setting.
+/PRINT/HI or
or
"is
"
9