Decagon Devices DS-2 Manual del usuario - Página 18

Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida Decagon Devices DS-2. Decagon Devices DS-2 20 páginas. Sonic anemometer

DS-2
6

Troubleshooting

Problem #1: Sensor not responding:
Check power to the sensor
Check sensor cable and 3.5 mm plug integrity
Check data logger wiring to ensure
White - 3.6 to 15 V supply
Red - Digital out
Bare - Ground
If sensor does not respond, use the Pro-Check to make sure it is
working satisfactorily.
6 TROUBLESHOOTING
15