Major tech MT943 Manual de instrucciones - Página 5
Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida Major tech MT943. Major tech MT943 12 páginas. Data logging light meter
3.SPECIFICATIONS
Function
Display
Measuring Range
Over Range Display
Spectral Response
Spectral Accuracy
Cosine Response
Accuracy
Repeatability
Sampling Rate
Photo Detector
Operating Temperature 0 to 40°C (32 to 104°F)
Operating Humidity
Storage Temperature
Storage Humidity
Power Source
Photo Detector
Lead Length
Photo Detector
Dimensions
Meter Dimensions
Weight
Accessories
Range
3-3/4 digit LCD with high speed 40 segment bar
graph.
400.0Lux, 4000Lux, 40.00kLux and 400.0kLux /
40.00FC, 400.0FC, 4000FC, 40.00kFC
Note: 1FC=10.76Lux, 1kLux=1000Lux,
1kFC=1000FC
LCD will show "OL" symbol.
CIE Photopic (CIE human eye response curve).
CIE Vλ function f1' ≤6%
f2' ≤2%
±3% rdg±0.5%f.s. (<10,000Lux); ±4%
rdg±10d. (>10,000Lux)
±3%
1.3 times/sec of analog bar-graph indication;
1.3times/sec of digital display. Datalogger
sampling could be setup.
One silicon photo diode and spectral response
filter.
0% to 80%RH
-10 to 50°C (14 to 140°F)
0% to 70% RH
1 piece 9V battery
150cm (approx.)
115 x 60 x 20mm (L x W x H)
170 x 80 x 40mm (L x W x H)
390g
Carry Case, Instruction Manual, Battery
5