Gamry Interface 5000 Instrucciones

Navegue en línea o descargue pdf Instrucciones para Equipos de laboratorio Gamry Interface 5000. Gamry Interface 5000 2 páginas. Installing standard cable on reference 600+
También para Gamry Interface 5000: Manual de inicio rápido (4 páginas)

Gamry Interface 5000 Instrucciones
Instructions for installing Standard Cable
on Reference 600+
Hardware Deliverables
Calibration Cell (Interface 5000 Calibration Cell/200 Ohm)
60cm Standard Cell Cable (PN 985-00071)
Software Deliverables
IE Correction file (REF600P_CID14.IECOR)
Software Setup
Install the Gamry software as described in the normal documentation. Copy the REF600P_CID14.IECOR
to the following location: C:\ProgramData\Gamry Instruments\Framework\IE Corrections. This folder
may be hidden on some computer, but pasting the above italicized name into Explorer should make it
visible. Please note that for international versions of Windows, the ProgramData folder may be named
slightly different.
Calibration Setup
Connect the 25 pin male end of the 60cm cell cable to the 25 pin female connector on the front of the
Reference 600+. Connect the cell leads of the 60cm cable to the Calibration cell as shown in Figure 1.
Figure 1: Calibration cell showing proper connections.
1