4

Operation

NOTICE
When the CAPBs is switched on, the measuring instrument performs
zero calibration.
5

Maintenance

6
Storage
temperature
Humidity
Dimensions
W x H x D [mm]
CAPBs STm
hours of
operation
Application
When the CAPBs is switched on, the measuring instrument
performs zero calibration. Zero calibration must be performed in
fresh air with a CO2 concentration of approx. 400 ppm and a TVOC
concentration of approx. 125 ppb.
INCORRECT MEASURED VALUES
If you want to take a measurement, verify that you only switch
on the measuring instrument in fresh air with a CO2
concentration of approx. 400 ppm and a TVOC concentration
of approx. 125 ppb.
Failure to follow these instructions can result in equipment damage.
The measuring instrument does not contain any parts that can be
serviced or repaired by the customer.
Prior to performing a measurement, perform a visual inspection of
the CAPBs for visible damage. Do not use damaged measuring
instruments.
-20 °C to +60 °C
0 ... 95 % rh, non-condensing
58 x 42 x 35
Up to 60 hours of battery operation
Detection of air quality
in building technology
®
CAPBs