decon INFM7202 Manual de instrucciones de montaje - Página 5

Navegue en línea o descargue pdf Manual de instrucciones de montaje para Silla de ruedas decon INFM7202. decon INFM7202 8 páginas. Infinity mounting set panthera s3/u3/u3 light/u3 light l

Mounting set
INFM7002, INFM7102, INFM7202
__________________________________________________________________________________
Now mount the complete
adapter bar that consists of
center pipe (A), outer pipe (B)
and clamp (C). The complete
adapter bar is mounted into
the holders (D) by using the
QR-pins (E).
Make sure the ends of the
outer pipe (B) are parallell
otherwise it can be hard to
assemble it into the holders
(D). It shall be possible to move
the center pipe (A) sideways
and also rotate it if correctly assembled.
First tighten the screws in the
panthera clamp (H), it shall be
possible to move the center pipe
sideways and also rotate it even
after tightening of the screws. Now
tighten the screws in the universal
clamp (G).
INFO
Check that the arrow on the labeling mounted on the center pipe (A) is pointing towards the
front of the wheelchair. If the center pipe (A) is mounted the wrong way around it will not be
possible to attach Infinity to the bar.
WARNING!
The screws holding the panthera clamp (H) shall be tightened to 15-20Nm of torque.
The screws holding the universal clamp (G) shall be tightened to 10-15Nm of torque.
__________________________________________________________________________________
Page 5/8