5. Error messages
EXPLANATION OF THE DISPLAY MESSAGES
Error
Description
E13
Lamp start fail
E14
Over-
(TEMP)
temperature
E15
Fan fail
(FAN)
E17
Temperature
sensor
U51
Distance
(DIST)
U94
Time
(TIME)
management
Solution
Turn both switches OFF.
Start from 4 PUTTING INTO OPERATION.
In case of recurrence E13, please consult DecoRad®
Systems
The UV lamp is too hot, leave the UV lamp to cool down
for 5 minutes. Turn both switches OFF.
Start from 4 PUTTING INTO OPERATION.
In case of recurrence, please consult DecoRad® Systems
Leave the UV lamp to cool down for 5 minutes. Turn both
switches OFF.
Start from 4 PUTTING INTO OPERATION.
In case of recurrence, please consult DecoRad® Systems
Leave the UV lamp to cool down for 5 minutes. Turn both
switches OFF.
Start from 4 PUTTING INTO OPERATION.
In case of recurrence, please consult DecoRad® Systems
The lamp-unit is turned too far from the object for more
than 30 seconds.
Please position the Handmate between 2 – 4 cm distance
from the floor, cure at approx. 3 cm
Auto OFF after 5 minutes no use
Handmate UV 1800 Firmware 2.9
15