Sleep Timer Setting [ SLEEP ]
1. Sleep Timer settings can also be configured in the Acaia apps.
2. The number indicates the amount of time before the scale automatically powers off. For example,
when 10 is selected, the scale powers off after 10 minutes of being idle if the scale is disconnected
from all mobile devices.
Zero Tracking Setting [ trAcE ]
1. Zero Tracking is a function for automatically maintaining the zero indication within certain limits.
2. OFF: It is not recommended to turn off Zero Tracking because the zero indication can drift easily
as the room temperature changes.
3. 0.5 d~3 d: For example, for standard resolution each division is 0.1 g, 3 d = 0.3 g. When Zero
Tracking is set to be 3 d, any initial weight between +/- 0.3 g on the weighing pan will be zeroed out
gradually.
4. 2 d is recommended in general for the Acaia Lunar.
Key Sound Setting [ bEEP ]
1. Key Sound settings can also be configured in the Acaia apps.
2. Set the Key Sound function to on/off to enable/disable the beeping sound.
Zero Range Setting [ 0.rAnGE ]
1. The Zero Range setting determines the range for which the tare button executes zero operation
instead of tare operation.
2. When Zero Range is set to 2, the range for zero-setting will be 2% of the capacity, i.e., +/- 40 g
for Acaia Lunar. By pressing the tare button, any weight within +/- 40 g is zeroed out by zero opera-
tion while any weight that lies outside the range is tared out by tare operation.
3. If Zero Range is set to 100, the range for zero-setting will be 100% of the capacity, i.e., +/- 2 kg
for Acaia Lunar. Therefore, instead of tare operation, zero operation would be executed in this set-
ting whenever tare button is pressed.
Notice: The difference between zero operation and tare operation is that zero operation would sim-
ply ignore the weight and display as 0 while tare operation would deduct the weight from the weigh-
ing result. However, zero operation does not increase the scale capacity.
Weighing Filter Setting [ Filt ]
1. The Weighing Filter Setting determines the speed at which the load receptor responds to weight
and determines the weighing value. The larger the number, the more stable but slower the weighing
result will be. The number shown on the LED display indicates the current filter setting.
2. Press tare button to change the parameter.
Default Unit Setting [ Unit ]
1. The Default Weighing Unit is set to grams if [ 9 ] is selected and ounce if [ o2 ] is selected.
Weighing Resolution Setting [ rESoL ]
1. The default Weighing Resolution is set to [ dEFAUL ]: the division is 0.1 g/0.002 oz.
2. Tap tare button to switch the Weighing Resolution between [ dEFAUL ] and [ hiGh ].
3. When the Weighing Resolution is set to high, division will be 0.01 g/0.0005 oz. The default display
will show [ 0.00 ] when the scale is set to grams, and [ 0.0000 ] for ounces in Weighing Mode.
Note: After setting Weighing Resolution to high, the default display will show [ 0:00 0.00 ] with timer
in Dual Display Mode and Auto Start Modes.