Allied Healthcare Schuco S132 Manual de inicio rápido - Página 5

Navegue en línea o descargue pdf Manual de inicio rápido para Equipamiento médico Allied Healthcare Schuco S132. Allied Healthcare Schuco S132 12 páginas.

READ AND SAVE THESE INSTRUCTIONS
SET-UP:
1.
In order to become familiar with this unit, note where the following features are: ON/OFF switch, vacuum gauge, control knob and
hydrophobic bacterial filter.
2.
The mechanical shut-off valve under the lid will stop suction if collected fluids reach the shut-off valve level. If this happens, turn the unit off
and empty the bottle or replace disposable canister.
3.
This product has been provided with a hydrophobic filter, which will protect the pump from contaminants. WHEN THIS FILTER COMES IN
CONTACT WITH LIQUID, THE SUCTION FLOW WILL STOP (MOTOR WILL STILL RUN). IF THIS HAPPENS, THE FILTER MUST BE
REPLACED. It is suggested that the filter be replaced after each patient use or on a monthly basis if the same patient uses the unit or
when a reduction in airflow is detected. To replace the filter, remove the hose from the hose barb on the back of the filter. Unscrew the
filter from the control valve and discard. Carefully screw the fresh filter into the control valve and reconnect the hose to the hose barb. Do
not over tighten filter because this may cause damage to the pump. Replacement filters (Part Number 01-90-3100 or 01-90-3100-KIT)
should be kept on hand.
VACUUM GAUGE ASSY.
ON/OFF SWITCH
|
W/CONTROL KNOB AND FILTER INPUT
HYDROPHOBIC BACTERIA FILTER
REMOVE PLUG AND SCREW FILTER INTO REGULATOR BODY