TROUBLE SHOOTING
Bath Does not move
Check power.
Check fused spur to ensure there is power to the unit.
Ensure the RCCB is working and has power to it. To test the RCCB push the test button; the
RCCB should interrupt power and the switch should trip. Push the RESET button; power should be
restored. If the RCCB fails to operate in this manner there is a problem with the electric supply
and the service department should be contacted.
If mains power is interrupted the bath will only work on battery backup. The bath will continue to
operate but as the batteries become low the bath will emit an audible beep when the handset is used.
Once the warning beep starts, lower the bath and do not use until the fault is rectified. Once power is
restored allow the batteries to recharge before operation. (Usually 3 hours).
If the handset buttons do not work, but do however emit a short beep when pressed, the handset may
be locked.
To unlock/ lock the handset see 'Handset Lock' section.
Bath judders.
Usually occurs if the bath has not been fixed to the floor correctly or the bath is not level.
Bath noisy
Check to ensure there is nothing touching the bath while it moves
Ensure the pipe-work and cables are free of the panels and not rubbing against the framework
Check panels are free to move and are not touching each other.
Bath Leaking
Look for source of leak – typically caught flexi-hose on feed or waste. This can usually be
remedied by your own maintenance contact.
Bath is uneven (one end higher than the other)
This may occur after a power failure. Use the DOWN button on the handset. Lower the bath to
its lowest position. This will automatically reset the bath so that both sides rise together.
Bath Beeping
Mains power has failed to the unit – See section titled 'Bath does not move'
Bath Panels touching /noisy
If the upper and lower panels touch, marking will occur. Adjust panels as necessary to correctly
align.
Gemini 2000 User Manual v1.3
Some basic checks are detailed below.
If in doubt contact service support on 01527 400024.