Abaxis VetScan HM5 Manual de consulta rápida - Página 13
Navegue en línea o descargue pdf Manual de consulta rápida para Instrumentos de medida Abaxis VetScan HM5. Abaxis VetScan HM5 19 páginas. Hematology analyzer
También para Abaxis VetScan HM5: Manual de consulta rápida (17 páginas), Manual de instalación (1 páginas), Instrucciones de instalación (2 páginas)
Blank Warning Flags
Flag
Meaning
p
PLT blank is high
X
EOS blank high
b
RBC blank is high
More thorough troubleshooting procedures can be found in the Abaxis HM5 Operator's Manual.
Warning Flags
Visually inspect the wash head and clean if necessary; run 2-3
automatic self cleanings, then repeat the blank. If the warning
persists, run Soak Cleaning and the repeat the Blank.
Visually inspect the wash head and clean if necessary: check
Diluent (green) and Lyse2 (orange) lines for loose connections;
prime Diluent and/or Lyse2 if bubbles are present or were
introduced while checking the tubing; repeat the blank.
Visually inspect the wash head and chamber below it for
blood debris; clean if necessary. Run a Soak Cleaning, then an
automatic self-cleaning; repeat the Blank.
(Partial List)
Solution