7. Draining the water tanks

1.
Connect the supplied drain hose to
the drain valve by pressing it on firmly.
3.
Pull the drain valve outward, the
tank will begin to drain through the hose.
The sterilizer includes two draining valves. The left valve is used
for draining the used water tank. The right valve is used for draining
the clean water tank. Always be sure that the hose is draining into
a vessel with at least 2L capacity.
9
2.
Set the drain valve to the open
position by turning it counter clockwise
4.
After you have completed draining the
water, push the drain valve inward and
set to the closed position
Instructions manual
10.5 Replacement of the door seal
10.5.1 Fully open the door.
10.5.2 Remove the door seal carefully by pulling it by hand away from
the door
10.5.3 Clean the new door seal carefully with a soft cloth saturated
with distilled water.
10.5.4 Moisten the new seal with medical disinfectant .
10.5.5 Insert the new seal following the instruction below:
1). Press on the top of the seal.
3). Press on the left and right
portions of the seal.
Caution: Always ensure the chamber and the door have
been cooled down before attempting to change the seal.
Instructions manual
2). Press on the bottom of the
seal.
4). Press the remaining seal surface
and ensure that it is smooth.