Dickson KT6P2 Primeros pasos - Página 6

Navegue en línea o descargue pdf Primeros pasos para Grabadora de voz Dickson KT6P2. Dickson KT6P2 7 páginas. Remote sensing temperature recorder
También para Dickson KT6P2: Manuales de inicio rápido (9 páginas)

Why does the display read PROB?
Why does calibration seem to be off?
Why won't the battery back-up work?
Why won't the chart recorder respond to changes?
Why won't the unit power up?

WARRANTY

Dickson warrants that this line of instruments will be free from defects in material and
workmanship under normal use and service for a period of twelve months after delivery.
This warranty does not cover routine calibration and battery replacement.
Dickson KT6P Manual V3
1.800.561.8187
Check the sensor pod and KTC probe(s). Make sure it is properly plugged into the unit.
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit
may appear to be working, but the readings won't change, also the chart will not rotate.
Remove power and battery, and then re-power.
What is the tolerance of the unit it's being compared to?
It's ok if the unit is within the sum of the two tolerances.
Has an outside calibration house attempted calibration? It may not have been adjusted
properly.
Walk through calibration adjustment, found in the manual
Are there good batteries in the Chart Recorder?
Keep in mind that battery backup varies significantly depending on temperature, pen
movement and chart rotation speed.
If there was a momentary power outage (Brown out), the unit may not have had enough
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or
shut off in this situation) This situation may also occur if the unit is plugged into an outlet
that shares the circuit with other machinery that has phase motors or compressors that
cycle periodically. As these other machines cycle, they momentarily have a high current
draw, therefore drawing from the unit.
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit
may appear to be working, but the readings won't change, also the chart will not rotate.
Remove power and batteries, and then re-power.
Remove the batteries and power adapter for a minute or two, this will reset the unit. The
unit should respond when the adapter is plugged in.
www.
6
.com