OPERA TION
The system is designed to operate continuously, with a minimum amount of operator
input.
It is, however, advisable to inspect the sample loop periodically to ensure that the
required pressures and flows are being maintained.
The number and type of items employed in the sample loop will determine what, if any,
other routine checks should be made. If, for instance, a filter is used, the filter element
should be inspected periodically and changed when necessary.
The instrument should not require any routine maintenance but if any malfunction is
suspected it is advisable to contact your local dealer.
Should it be necessary, at any time or for whatever reason, to change either the
instrument or sensor, it should be noted that the components of the DS 1000 system are
fully and completely interchangeable provided that the conesponding instrument/sensor
range is requested. The only adjustment necessary would be the alarm set points in the
case of the instrument.
While the sensor should give several years operation, it is advisable to confirm the
calibration, from time to time, to ensure accurate operation of the system.
The sensor is equipped with an 'Autocal'
feature which allows adjustment of the
calibration span. To operate the' Autocal' use the following procedure:-
1.
Refer to sensor assembly drawing page 11. Unscrew the black serrated retaining
ring from the cable plug and carefully lift the plug away from the sensor body, to
expose the internal PCB. Reconnect the sensor cable.
2.
Locate the small potentiometer on the top edge of the PCB.
3.
Expose the sensor to a known dewpoint (e.g. ambient air, or a calibration gas) and
adjust the' Autocal' potentiometer until the display reads the known dewpoint.
4.
Re-assemble the sensor plug, ensuring that the '0' ring is cOlTectlypositioned and
re-install the sensor in its nOlmal operating position.
5.
This completes the procedure