General Cleaning Guidelines
Gently wipe the eVER-HOME and accessories as needed with a lint-free microfiber cloth dampened with a
small amount of distilled water. Avoid getting excess moisture onto the devices; the corner of the microfiber
cloth is sufficient. You may also use a camera lens cleaning cloth.
Make sure the devices are powered-down before cleaning. Remove the eVER-HOME from the protective
cover.
DO NOT
apply too much pressure.
DO NOT
use compressed air, as it may damage the devices.
Disinfecting Guidelines
Gently wipe the devices using a disinfectant containing hypochlorous acid-based solution containing
50–80ppm (such as Clorox or Lysol disinfectant wipes), or an alcohol-based solution containing more than
70% ethanol or isopropyl alcohol. Do not apply these liquid solutions directly to the eVER-HOME; they
should be carefully applied with a pre-moistened wipe or by dampening the corner of a microfiber cloth with
a small amount of the disinfectant.
Disinfecting equipment
1. Mask and gloves are recommended.
2. Remove all contents from the Carrying Case.
3. Remove the tablet from the Tablet Cover.
4. Tablet: Wipe the tablet on all sides using the guidelines above. Inspect the device front, back, and ports
and buttons for any damage or foreign debris.
5. Blood Pressure Monitor: Wipe the monitor on all sides using the guidelines above. Inspect the device
front, back, and all sides, ports and buttons for any damage or foreign debris. Wipe the cuff, hoses, and
connectors using the guidelines above. Inspect the hoses and connectors for any damage or foreign
debris. Inspect the cuff Velcro for foreign debris. Make sure the area where the hose connects to the cuff
is unbroken.
6. SpO2 Sensor: Wipe the SpO2 sensor on all sides and inside the finger opening using the guidelines above.
Be sure to reach the hinge area inside. Inspect the device front, back, and all sides, joints and crevices for
any damage or foreign debris.
7. Thermometer: If still attached, eject and dispose of the probe cover. Throw away all unused probe covers.
Wipe the thermometer on all sides and around the probe area using the guidelines above. Inspect the
device front, back, and all sides, joints and crevices for any damage or foreign debris.
8. Scale: Wipe the scale on all sides using the guidelines above, paying attention to crevices and cracks.
Inspect the device front, back, and all sides, joints and crevices for any damage or foreign debris.
9. Carrying Case: Remove dividers. Clean all surfaces with a mild cleaning solution, and wipe down
according to the guidelines above.
10. Tablet Cover: With tablet removed, clean all surfaces and around the corner clips with a mild cleaning
solution and wipe down according to the guidelines above.
DO NOT
use household cleaners as they may damage the devices.