Olympus 1X71 Manual del usuario - Página 9

Navegue en línea o descargue pdf Manual del usuario para Microscopio Olympus 1X71. Olympus 1X71 18 páginas.

Adjusting the objective correction collar

Correction is possible according to the vessel bottom thickness.
1. When the thickness of the vessel bottom is known, match the scale reading of the
correction collar to the thickness of the vessel in use.
or
2. If the thickness of the vessel is unknown or diverge from the manufacturer
specifications, the optimum position for the correction collar can be obtained by
judging the image resolution and contrast. When a satisfactory image is not obtain
after focusing:
1. Rotate the correction collar to the left and right, refocus each time and
compare the images.
2 Then rotate the collar in the direction yielding a better image, rotate the
correction collar to the left and right, refocus each time and compare the
images.
3 Repeat this cycle until the position with the optimum image is found.
20x Correction Collar
Correction Collar Scale
0 mm
0.17 mm (glass coverslip #1.5)
0.5 mm
1 mm
(most tissue culture plates)
1.5 mm
2 mm
40x Correction Collar
9