Olympus BH-2 BHT Manual - Página 17

Navegue en línea o descargue pdf Manual para Microscopio Olympus BH-2 BHT. Olympus BH-2 BHT 24 páginas. Sliding focus block
También para Olympus BH-2 BHT: Desmontaje, limpieza y montaje completos (23 páginas)

Olympus BH-2 BHT Manual
Lamp House Repair Procedure
As an alternative to the J.C. Ritchey repair kit, a Bender and Wirth 990 lamp socket (available from J & H Microscope 
Services  or  from  Bulbworks)  may  be  used  to  replace  the  existing  lamp  socket.    To  replace  the  socket  with  the  new 
Bender & Wirth part, start by removing the halogen lamp and setting it aside, being careful to not touch the glass with 
your  fingers.    Next,  remove  the  lamp  reflector,  if  present,  by  removing  the  small  screw  securing  the  reflector  to  the 
support post.  Remove the back cover by carefully heating the perimeter of the housing with a heat gun to loosen the 
adhesive, while pushing from inside the housing, behind the lamp socket, with a screwdriver tip to free the loosened 
back  piece.    Avoid  direct  heat  or  pressure  to  the  ventilation  slots  on  the  back  cover,  as  they  are  fragile  and  can  be 
damaged.    Once  the  back  cover  has  been  removed,  clean  off  any  remaining  adhesive  and  remove  the  two  screws 
securing the old lamp socket in place.  Unsolder the two wires to free the old socket from the housing.  Trim and strip 
the wires on the new socket to match those of the original socket, and solder the new socket in place.  Reinstall the two 
screws  to  secure  the  new  socket  in  its  proper  position.    Reinstall  the  back  cover,  securing  with  suitable  epoxy,  and 
reinstall the lamp reflector (if applicable).  Reinstall the halogen lamp (better still, install a new lamp), being careful to 
not touch the glass with your fingers.  If the glass is accidentally touched, clean it with isopropyl alcohol before use. 
Halogen Lamp 
The BHT/BHTU stand uses a Philips #7388 halogen projector lamp with a clear quartz envelope and a G4 base (see Figure 
20).    This  lamp,  which  has  a  color  temperature  of  3200K,  a  color  rendering  index  of  100,  and  outputs  475  lumens,  is 
rated  for  6V/20W  with  an  advertised  life  of  100  hours  at  full  brightness.    The  Philips  #7388  may  be  purchased  from 
Bulbworks  or  Amazon.    Alternatively,  any  of  the  manufacturer's  part  numbers  listed  in 
replacements as well. 
 
                                                             
1
 Lamp house repair procedure courtesy Jerry Clement of J & H Microscope Service, Inc. 
Olympus BH‐2 (BHT/BHTU) Electronics 
1
 
Figure 20 –Philips #7388 FHE/ESB T3 Halogen Projection Lamp 
Manufacturer 
Philips 
Osram 
Ushio 
GE 
Nikon 
Olympus 
Reichert 
Swift 
Table 1 – Suitable Halogen Lamps for Olympus BHT/BHTU Microscopes 
 
 
 
778 (49718) or 788 (943117) 
 
   Revision 2 
Table 
 
Part Number 
7388 (256784) 
54261 or 64250 HLX 
FHE/ESB (1000532) 
79099 
8‐C405 
11143 
MA‐780 
   
 
 
1  may  be  used  for 
        Page 17 of 24