Olympus BX3-SSU Manual de instrucciones - Página 10
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus BX3-SSU. Olympus BX3-SSU 24 páginas. Scanning stage with ultrasonic
También para Olympus BX3-SSU: Manual de instrucciones (20 páginas)
4
4
This stage is equipped with the operation parameters to provide the motorized operation.
If the stage operation parameters are displaced against your use environment, you may not obtain the best operational
feeling, such that the stage does not move smoothly, etc. In that case, compensate the stage operation parameters.
4-1 When Using BX63
1
2
7
Downloaded from
ManualsNet.com
COMPENSATION OF THE STAGE
Fig. 5
Fig. 6
search engine
When using BX63, the auto compensation of the stage can be performed
from the touch panel controller.
Before starting the auto compensation, install the stage on the
CAUTION
BX63 and make sure nothing, including a specimen, is placed
on the stage.
1. Open the [Menu] display by controlling the touch panel.
2. Tap the System Setting button @ to open the [System Setting] display.
3. Tap the [ECO, Others] tab 2 on the [System Setting] display.
4. Tap the [Setting 2] tab 3 and then tap the Auto Compensation button
4.
5. When the auto compensation confirmation display appears, tap [OK] to
start auto compensation.
The stage moves automatically during the auto compensation
CAUTION
(for a few minutes). Do not touch the stage during the movement.
3
6. When auto compensation completes, the touch panel screen shows the
completion message.
If the foreign substance is entered in the stage driving part,
CAUTION
the status may not be recovered by performing the Auto
4
Compensation. In this case, turn the power of BX3-SSU Control
Box to "
" (OFF), and perform the Auto Compensation again
after moving the stage from front to back and from side to side
several times with a hand.