Olympus BX41-P Manual de instrucciones - Página 26
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus BX41-P. Olympus BX41-P 36 páginas. Polarizing microscope
También para Olympus BX41-P: Manual de instrucciones (36 páginas)
Fig. 31
23
3
Adjusting the Eyepiece Cross Lines
}To align the eyepiece cross lines and the vibration direction, the
orientation plate provided with the U-CPA or U-OPA is required.
Orientation plate (U-PJ: Single module name)
CAUTION
Remove the test plate adapter and the compensator from the
light path.
1. Swing out the top lens of the condenser.
2. Place the orientation plate on the stage. Remove the analyzer from the
light path, and use the 4X objective to focus.
(Fig. 31)