Olympus BXC-CBB Manual de instrucciones - Página 15
Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus BXC-CBB. Olympus BXC-CBB 41 páginas. Modular microscope assemblies hardware
TIP:
1
*
AS control
The AS mechanism uses backlash. For this reason, the AS diameter after moving varies depending whether the diaphragm
was opened or closed. This is referred as lost motion. The lost motion for this system corresponds approximately to 200
pulses. This phenomenon is more apparent when the AS diameter is smaller. The following figure shows the relationship of
the AS diameter to the specified pulse numbers using the 3-stage action, ① ⇒ ② ⇒ ③ , as an example.
Normal
In the normal mode, the specified AS diameter is achieved with the shortest path. As action ③ includes a lost motion, the
AS diameter defers from the AS diameter when stopped at action ① . Based on the recommended pulse number (see the
table on page 28) for each objective, select high-precision control, if necessary.
High-precision control
By selecting high-precision control, you can reduce the effect of the lost motion. In this mode, the positioning accuracy
improves as the system always stops at the specified AS diameter from the opening direction. However, if the opening
direction is specified, the control time will be longer for the amount of time it takes to move the distance shown with the
red arrow.
2
*
NP Control
Selecting "Rotate through the highest numbered hole" rotates the nosepiece to the specified hole using the shortest path.
Selecting "Does not rotate through the highest numbered hole" rotates the nosepiece in the opposite direction if the
shortest path to the specified hole goes through the highest numbered hole. By attaching a low-magnification objective to
hole 1 and a high-magnification objective to the highest numbered hole, you can avoid an objective colliding with the
sample when the nosepiece is rotated.
11
3. Setup procedures
3.5 Setting up BXC-CBB