Olympus BXC-FSU Manual de instrucciones - Página 14

Navegue en línea o descargue pdf Manual de instrucciones para Microscopio Olympus BXC-FSU. Olympus BXC-FSU 41 páginas. Modular microscope assemblies hardware

3.5 Setting up BXC-CBB

TIP:
Before setting the DIP switch, set the main switch to ○(OFF).
The switch settings are read and defined after Power-ON.
1.
Confirm that the DIP switch on BXC-CBB is set properly.
The allocated functions of the DIP switches are shown in the table below.
: Factory default setting
SW-No.
1
2
3
4
5
Off
On
Off
On
Off
On
Off
On
Off
On
3. Setup procedures
Function
6
7
8
Buzzer sound
Nosepiece (NP)
1
AS control*
NP Control*
Reserved by maker Always OFF
Off
Reserved by maker Always OFF
On
Off
Reserved by maker Always OFF
On
Off Reserved by maker Always OFF
On
The buzzer sound is heard
The buzzer sound is not heard
5Position-NP
6Position-NP
Normal
High-precision control (lost motion reduction)
2
Rotate through the highest numbered hole
Does not rotate through the highest numbered hole
Note
3.5 Setting up BXC-CBB
10